Refractive topographic maps in assessing the quality of the ocular optical system with complex-design IOLs
https://doi.org/10.29413/ABS.2026-11.4.16
Abstract
Objective assessment of postoperative eye refraction in patients with implanted intraocular lenses (IOLs) featuring addition zones or a modified wavefront profile is challenging using conventional methods.
Aim. To analyze the diagnostic value of the method of objective refractive topographic mapping of the eye in patients with IOLs of complex optical design.
Material and Methods. A total of 198 patients (mean age 66.4 ± 7.8 years) with various types of IOLs – monofocal, multifocal, and extended depth of focus lenses were initially screened. Inclusion criteria: the absence of pathology of the cornea, vitreous, retina, or optic nerve; a postoperative follow-up period of six months or more. Exclusion criteria: uncorrected distance visual acuity of 0.7 or worse, opacity or folds of the posterior lens capsule. After applying inclusion and exclusion criteria and performing random eye selection (right or left) for eligible participants, 192 eyes were included in the final analysis. One eye per subject was chosen at random to avoid inter-eye correlation. All patients underwent a comprehensive ophthalmological examination, including objective assessment, visometry, tonometry, depth of focus (DoF) calculation based on the defocus curve (the diopter range within which visual acuity remains no lower than 0.5), wavefront aberrometry (Peramis, SCHWIND eye-tech-solutions GmbH, Germany) with refractive topographic map (RTM) construction and calculation of the refractive inhomogeneity index.
Results. The construction of RTMs revealed the distribution of refractive indicators across the entire scanning area and the presence of distinct addition zones in eyes with multifocal IOLs and extended depth of focus IOLs. Significant differences were found in the refractive inhomogeneity index values between groups (H(7) = 46.24, p = 0.0001) and varying levels of its consistency with DoF.
Conclusion. The method of pyramidal sensor-based aberrometry with refractive topographic mapping represents a highly informative approach for the objective assessment of the eye’s optical system in patients with complex-design IOLs and could serve as a foundation for the development of personalized cataract surgery.
About the Authors
O. I. RozanovaRussian Federation
Olga I. Rozanova – Dr. Sc. (Med.), ophthalmologist, head of the Scientific and Educational Department of Irkutsk Branch of S. Fyodorov Eye Microsurgery Federal State Institution; assistant of the Department of Ophthalmology, Irkutsk State Medical Academy of Postgraduate Education – branch of the Russian Medical Academy of Continuing Education professional Education of the Ministry of Health of the Russian Federation
Lermontov str., 337, Irkutsk 664033,
Yubileinii dstr., 100, Irkutsk 664049
B. A. Tsybzhitova
Russian Federation
Balzhinima A. Tsybzhitova – ophthalmologist at the 1st Ophthalmological Department
Lermontov str., 337, Irkutsk 664033
References
1. Kohnen T, Berdahl JP, Hong X, Bala C. The Novel Optical Design and Clinical Classification of a Wavefront-Shaping Presbyopia-Correcting Intraocular Lens. Clin Ophthalmol. 2023; 17: 2449–2457. doi: 10.2147/OPTH.S400083
2. Morozova TA. Extended Depth-of-Focus is a New Intraocular Lens Technology in the Treatment of Presbyopia: Review. Bulletin of the Russian Academy of Medical Sciences. 2024; 79(4): 318–326. (In Russ.). doi: 10.15690/vramn17915
3. Hoffer KJ, Savini G. Update on intraocular lens power calculation study protocols. Ophthalmology. 2021; 128(11): e115–e120. doi: 10.1016/j.ophtha.2020.07.005
4. Ilyina NV. Aberrations and other factors that determine the functional result of phacoemulsification of cataract. Practical medicine. 2012; 4(59): 260-262. (In Russ.).
5. Ophthalmology: national guide / ed. SE Avetisova, EA Egorova, LK Moshetova, VV Neroeva, KhP Takhchidi. – 3rd ed., revised. and additional – Moscow: GEOTAR-Media, 2024. (In Russ.).
6. Coassin M, Di Zazzo A, Antonini M, Gaudenzi D, Gallo Afflitto G, Kohnen T. Extended depth-of-focus intraocular lenses: power calculation and outcomes. J Cataract Refract Surg. 2020; 46(11): 1554–1560. doi: 10.1097/j.jcrs.0000000000000293
7. Ramazanova LSh, Napylova OA, Shamratov RZ. About the Possible Mechanisms of the Influence of Optical Intraocular Implants on the Characteristics of Neurosensory Rehabilitation and Adaptation in Patients with Pseudophakia. Literature Review. Ophthalmology. 2021; 18(1): 30–35. (In Russ.). doi: 10.18008/1816-5095-2021-1-30-35
8. Zhang L, Lin D, Wang Y, Chen W, Xiao W, Xiang Y, et al. Comparison of Visual Neuroadaptations After Multifocal and Monofocal Intraocular Lens Implantation. Front Neurosci. 2021; 15: 648863. doi: 10.3389/fnins.2021.648863
9. Wanten JC, Bauer NJC, Berendschot Tos TJM, van den Biggelaar FJHM, Nuijts Rudy MMA. Dissatisfaction after implantation of EDOF intraocular lenses. J Cataract Refract Surg. 2025; 51(5): 399–405. doi: 10.1097/j.jcrs.0000000000001615
10. Rudelevicius P, Lekaviciene R, Auffarth GU, et al. Relation between patient personality and patients’ dissatisfaction after multifocal intraocular lens implantation: Clinical study based on the five factor inventory personality evaluaton. Eye (Lond) 2020; 34(4): 717–7245. doi: 10.1038/s41433-019-0585-x
11. Takii M. Objective evaluation of the refractive properties of a diffractive multifocal IOL using a visible light-based autorefractometer prototype. Invest. Ophthalmol. Vis. Sci. 2025; 66(8): 51.
12. Borkenstein AF, Borkenstein E. Autorefraction Pitfalls in Extended Depth-of-Focus Pseudophakia: A Case Report. Cureus. 2025; 17(7): e87409. doi: 10.7759/cureus.87409
13. Behndig A, Montan P, Stenevi U, Kugelberg M, Zetterström C, Lundström M. Aiming for emmetropia after cataract surgery: Swedish National Cataract Register study. J Cataract Refract Surg. 2012; 38(7): 1181–1186. doi: 10.1016/j.jcrs.2012.02.035
14. Rozanova OI, Kuzmin SV, Mishchenko OP, et al. Pyramidal sensory wavefront aberrometry as a method for visualizing intraocular refractive errors in articulation. Bulletin of the Volgograd State Medical University. 2024; 1: 56–63. (In Russ.) doi: 10.19163/1994-9480-2024-21-1-56-63
15. Schmid R, Borkenstein AF. Enhanced Depth of Focus Intraocular Lenses: Through Focus Evaluation of Wavefront-Shaping versus Diffractive Optics. Biomed Hub. 2023; 8(1): 25–30. doi: 10.1159/000529234
16. Borkenstein AF, Borkenstein EM, Luedtke H, Schmid R. Optical Bench Analysis of 2 Depth of Focus Intraocular Lenses. Biomed Hub. 2021; 6(3): 77–85. doi: 10.1159/000519139
17. Temirov NN, Temirov NE. Visual acuity and clinical refraction following implantation of various multifocal intraocular lenses. Ophthalmology in Russia. 2015; 12(2): 37–42. (In Russ.). doi: 10.18008/1816-5095-2015-2-37-42
18. de Silva SR, Evans JR, Kirthi V, Ziaei M, Leyland M. Multifocal versus monofocal intraocular lenses after cataract extraction. Cochrane Database Syst Rev. 2016; 12(12): CD003169. doi: 10.1002/14651858.CD003169.pub4
19. Pershin KB, Pashinova NF, Tsygankov AYu, Miyovich OP. The outcomes of the implantation of bifocal and trifocal IOLs for presbyopia. Cataract and refractive surgery. 2016; 16 (2): 23-29. (In Russ.).
20. Orenburkina OI, Absalyamov MSh, Bikbulatova AA, Burkhanov YuK. Results of femto-assisted phacoemulsification of cataracts with implantation of lenses of new generation. Practical medicine. 2017; 9(110): 58–62. (In Russ.).
21. Pershin KB, Pashinova NF, Tsygankov AYu, Antonov EA, Kosova IV, Korneeva EA. Outcomes of extended depth of focus intraocular lenses implantation. Russian Annals of Ophthalmology. 2024; 140(2): 40–46. (In Russ.). doi: 10.17116/oftalma202414002140
22. Molebny VV, Panagopoulou SI, Molebny SV, Wakil YS, Pallikaris IG. Principles of ray tracing aberrometry. J Refract Surg. 2000; 16(5): S572–S575. doi: 10.3928/1081-597X-20000901-17
23. Mrochen M, Kaemmerer M, Mierdel P, Krinke HE, Seiler T. Principles of Tscherning aberrometry. J Refract Surg. 2000; 16(5): S570–S571. https://doi.org/10.3928/1081-597X-20000901-16
24. Rozema JJ, VanDyck DE, Tassignon MJ. Clinical comparison of 6 aberrometers. Part 1: technical specifications. J Cataract Refract Surg. 2005; 31(6): 1114–1127. doi: 10.1016/j.jcrs.2004.11.051
25. Rozema JJ, VanDyck DE, Tassignon MJ. Clinical comparison of 6 aberrometers. Part 2: statistical comparison in a test group. J Cataract Refract Surg. 2006; 32(1): 33–44. doi: 10.1016/j.jcrs.2004.11.052
26. Campbell CE. Wavefront measurements of diffractive and refractive multifocal intraocular lenses in an artificial eye. J Refract Surg. 2008; 24(3): 308–311. doi: 10.3928/1081597X-20080301-15
Review
For citations:
Rozanova O.I., Tsybzhitova B.A. Refractive topographic maps in assessing the quality of the ocular optical system with complex-design IOLs. Acta Biomedica Scientifica. 2026;11(4):157-165. (In Russ.) https://doi.org/10.29413/ABS.2026-11.4.16
JATS XML

.png)