Biocompatibility assessment of experimental prototypes of russian endovascular embolization coils made of stainless steel with polymer fibers

https://doi.org/10.29413/ABS.2026-11.4.24

Abstract

Endovascular embolization coils are a key tool for minimally invasive vessel occlusion; however, their widespread use in Russian practice is limited by high cost and dependence on imported supplies. The development of Russian analogues requires a comprehensive assessment of the biocompatibility of designs that combine a metal frame with polymer fibers.

Aim. To evaluate the biocompatibility of experimental prototypes of embolization coils made of stainless steel with polymer fibers (silk, poly(glycolide-co-lactide), polypropylene, polyethylene terephthalate, polyvinyl chloride with barium sulfate).

Materials and Methods. Coil samples were fabricated by mechanical winding of stainless steel wire followed by manual fixation of polymer fibers in the inter-coil spaces. Hemolytic activity was assessed by the degree of hemolysis of human red blood cells upon contact with sample extracts. Cytotoxicity of the experimental coil samples with fibers was evaluated using human fibroblasts of the FEC-16-2 line by direct (co-culture) and indirect (cultivation in material extracts) methods.

Results. All studied samples met standard biocompatibility criteria: the hemolysis level did not exceed acceptable values (less than 2 %), and no cytotoxic effect was detected. The lowest hemolysis values were recorded for coils with silk fibers (0.005 ± 0.004 %). Samples with poly(glycolide-co-lactide) showed a somewhat higher hemolysis level (1.184 ± 0.212 %), which may be attributed to hydrolytic degradation of the polymer with a local decrease in pH.

Conclusion. Silk, polypropylene, polyethylene terephthalate, and radiopaque fibers based on polyvinyl chloride with barium sulfate are promising materials for the development of Russian embolization coils, offering a combination of hemocompatibility and functionality. The use of poly(glycolide-co-lactide) requires additional evaluation of long-term safety. The obtained results justify further preclinical studies, including the assessment of thrombogenicity and stability of fiber fixation under simulated blood flow conditions.

About the Authors

E. V. Chepeleva
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Russian Federation

Elena V. Chepeleva – Cand. Sc. (Biol.), Senior Researcher at the Laboratory of Experimental Surgery and Morphology, Institute of Experimental Biology and Medicine

Rechkunovskaya Str., 15, Novosibirsk 630055



A. A. Dokuchaeva
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Russian Federation

Anna A. Dokuchaeva – junior researcher at the Laboratory of Experimental Surgery and Morphology, Institute of Experimental Biology and Medicine

Rechkunovskaya Str., 15, Novosibirsk 630055



A. A. Vaver
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Russian Federation

Andrey A. Vaver – junior researcher at the Laboratory of Bioprosthetics, Institute of Experimental Biology and Medicine

Rechkunovskaya Str., 15, Novosibirsk 630055



S. V. Vladimirov
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Russian Federation

Sergey V. Vladimirov – junior researcher at the Laboratory of Bioprosthetics, Institute of Experimental Biology and Medicine

Rechkunovskaya Str., 15, Novosibirsk 630055



V. P. Borodin
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Russian Federation

Vsevolod P. Borodin – engineer at the Laboratory of Bioprosthetics, Institute of Experimental Biology and Medicine

Rechkunovskaya Str., 15, Novosibirsk 630055



K. V. Kozyr
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Russian Federation

Konstantin V. Kozyr – junior researcher at the Laboratory of Bioprosthetics, Institute of Experimental Biology and Medicine

Rechkunovskaya Str., 15, Novosibirsk 630055



A. G. Badoian
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Russian Federation

Aram G. Badoian – Cand. Sc. (Med.), researcher at the Research Department of Endovascular Surgery, Institute of Circulation Pathology

Rechkunovskaya Str., 15, Novosibirsk 630055



A. A. Baranov
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Russian Federation

Aleksey A. Baranov – researcher at the Research Department of Endovascular Surgery, Institute of Circulation Pathology

Rechkunovskaya Str., 15, Novosibirsk 630055



M. A. Makhmudov
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Russian Federation

Mamurjon A. Makhmudov – endovascular surgeon

Rechkunovskaya Str., 15, Novosibirsk 630055



O. V. Krestyaninov
Meshalkin National Medical Research Center, Ministry of Health of the Russian Federation
Russian Federation

Oleg V. Krestyaninov – Dr.Sc. (Med.), professor, leading researcher at the Research Department of Endovascular Surgery, Institute of Circulation Pathology

Rechkunovskaya Str., 15, Novosibirsk 630055



References

1. Oka S, Kohno S, Arizono S, Onishi Y, Fumimoto M, Yoshida A, et al. Enhancing precision in vascular embolization: evaluating the effectiveness of the intentional early detachment technique with detachable coils in complex cases. CVIR Endovasc. 2024; 7(1): 40. doi: 10.1186/s42155-024-00453-7

2. Lim CJ, Shahin Y. Emerging Techniques and Treatment Outcomes in Pulmonary Arteriovenous Malformations Embolisation: A Narrative Review. J Clin Med. 2025; 14(23): 8455. doi: 10.3390/jcm14238455

3. Lauzier DC, Huguenard AL, Srienc AI, Cler SJ, Osbun JW, Chatterjee AR, et al. A review of technological innovations leading to modern endovascular brain aneurysm treatment. Front Neurol. 2023; 14: 1156887. doi: 10.3389/fneur.2023.1156887

4. Xiao N, Lewandowski RJ. Embolic agents: coils. Semin Intervent Radiol. 2022; 39(1): 113–118. doi: 10.1055/s0041-1740939

5. Abdikeev NM. Implementation of import substitution plans in high-tech sectors of the domestic industry in the face of external sanctions. Nauchnye trudy Vol’nogo ekonomicheskogo obshchestva Rossii. 2022; (3): 202–214. (In Russ.). doi: 10.38197/2072-2060-2022-235-3-202-214

6. Makarenko ED. Mechanisms for promoting Russian medical equipment on the domestic market. Finansovye rynki i banki. 2024; (3): 24–29. (In Russ.). doi: 10.24412/2658-3917-2024-3-24-29

7. Nakhod VI, Krivenko AN, Butkova TV. Market Review for Implantable Medical Devices. City Healthcare. 2024; 5(4): 209–216. (In Russ.). doi: 10.47619/2713-2617.zm.2024.v.5i4p1;209-116

8. GOST ISO 10993-12-2023. Medical devices. Biological evaluation of medical devices. Part 12. Sample preparation and reference materials. Moscow: Standartinform; 2023. 20 p. (In Russ.).

9. GOST ISO 10993-4-2020. Medical devices. Biological evaluation of medical devices. Part 4. Selection of tests for interactions with blood. Moscow: Standartinform; 2020. 28 p. (In Russ.).

10. Yu Y, Wang J. Recent advances and challenges in silk fibroin-based small-caliber vascular grafts: A review. Int J Biol Macromol. 2025; 322(Pt 3): 146930. doi: 10.1016/j.ijbiomac.2025.146930

11. Agapova OI. Silk Fibroin and Spidroin Bioengineering Constructions for Regenerative Medicine and Tissue Engineering (Review). Sovremennye tekhnologii v meditsine. 2017; 9(2): 190-206. (In Russ.). doi: 10.17691/stm2017.9.2.24

12. Taysi NM, Erten Taysi A, Ercal P, Sismanoglu S. Mechanical Performance of Nonabsorbable Monofilament Suture Materials Tied with Different Suturing Techniques Under Various Knot Configurations: An In Vitro Study. J Funct Biomater. 2025; 16(12): 428. doi: 10.3390/jfb16120428

13. Horakova J, Mikes P, Saman A, Svarcova T, Jencova V, Suchy T, et al. Comprehensive assessment of electrospun scaffolds hemocompatibility. Mater Sci Eng C Mater Biol Appl. 2018; 82: 330–335. doi: 10.1016/j.msec.2017.05.011

14. Kalinin RE, Suchkov IA, Mzhavanadze ND, Korotkova NV, Nikiforov AA, Surov IYu, et al. Comparison of cytotoxicity of synthetic vascular rostheses in vitro. I.P. Pavlov russian medical biological herald. 2020; 28(2): 183–192. (In Russ.). doi: 10.23888/PAVLOVJ2020282183-192

15. Polaskova M, Sedlacek T, Polasek Z, Filip P. Modification of Polyvinyl Chloride Composites for Radiographic Detection of Polyvinyl Chloride Retained Surgical Items. Polymers (Basel). 2023; 15(3): 587. doi: 10.3390/polym15030587

16. Omidian H, Wilson RL. PLGA Implants for Controlled Drug Delivery and Regenerative Medicine: Advances, Challenges, and Clinical Potential. Pharmaceuticals. 2025; 18(5): 631. doi: 10.3390/ph18050631


Review

For citations:


Chepeleva E.V., Dokuchaeva A.A., Vaver A.A., Vladimirov S.V., Borodin V.P., Kozyr K.V., Badoian A.G., Baranov A.A., Makhmudov M.A., Krestyaninov O.V. Biocompatibility assessment of experimental prototypes of russian endovascular embolization coils made of stainless steel with polymer fibers. Acta Biomedica Scientifica. 2026;11(4):240-250. (In Russ.) https://doi.org/10.29413/ABS.2026-11.4.24

Views: 13

JATS XML


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 2541-9420 (Print)
ISSN 2587-9596 (Online)