Experimental study of the strength characteristics of fixation of the symphysis pubis with an original plate for the reconstruction of the anterior pelvic semi-ring

https://doi.org/10.29413/ABS.2021-6.3.19

Abstract

The destabilization of bone fixators has led to repeated surgical interventions that increased the risk of migrations of metal fixators, infectious and inflammatory complications as well. The modern trends in hip bone injury surgery are related to developing and using of metal fixators, which are effective in acute injury; however, sometimes the condition of the patient and the technical support of hospitals don’t allow performing surgery in the acute period, and these metal fixators are ineffective for chronic damage. Consequently, until now, the problem of finding the optimal design of structures for fixing chronic damage of anterior pelvic semi-ring which will be able to exclude its destabilization, is still of current interest.

The purpose of the study: developing and experimental researching of durable features of original plate for reconstruction of the anterior pelvic semi-ring.

Materials and methods: In order to ensure stable fixation of chronic pelvic injuries, the original metal plates have been worked out. They are made individually in accordance with the anatomical and functional structures of the anterior pelvic semi-ring of the patient with using additive technologies. The study of the reliability of the different variants of ostiosynthesis of the anterior pelvic semi-ring with using well-known pelvic plates and a new original design was carried out. Stability tests for different plate fixing methods and mechanical strength of metallophyxators were carried out on a universal test machine of LFM-50kN series.

Results: one plate fixation in tensile test showed the lowest result – 0.341 kN, a low result of shear loads was received with the same object. The best result was shown by the polyaxial monolithic plate fixation in case of stretching – 0.51 kN at the shear loads – 0.591 kN. Necessary force applied to destabilizing of the metal structure while using a polyaxial monolithic plate was a half-higher than stretched, and a third higher than at shear loads, it demonstrates the benefits of using these metal fixators.

Conclusion: It is experimentally confirmed that the polyaxial monolithic plate provides the highest stability of fixing anterior pelvic semi-ring indicator in contrast with fixing of one or two plates.

About the Authors

A. F. Lazarev
Federal State Budgetary Institution National Medical Research Center of Traumatology-Orthopedics named after N. N. Priorov of the Ministry of Health of the Russian Federation
Russian Federation

 Professor, Doctor of Medical Sciences, Head of the Center for Acute Injury and Its Consequences

 Russian Federation, 127299, Moscow, st. Priorova 10 



E. I. Solod
Federal State Budgetary Institution National Medical Research Center of Traumatology-Orthopedics named after N. N. Priorov of the Ministry of Health of the Russian Federation
Russian Federation

 Associate Professor, Doctor of Medical Sciences, Leading Researcher of the Center for Acute Injury and Its Consequences 

Russian Federation, 127299, Moscow, st. Priorova 10 



Y. G. Gudushauri
Federal State Budgetary Institution National Medical Research Center of Traumatology-Orthopedics named after N. N. Priorov of the Ministry of Health of the Russian Federation
Russian Federation

 Doctor of Medical Sciences, traumatologist-orthopedist of the Center for Acute Injury and Its Consequences 

Russian Federation, 127299, Moscow, st. Priorova 10 



E. I. Kalinin
Federal State Budgetary Institution National Medical Research Center of Traumatology-Orthopedics named after N. N. Priorov of the Ministry of Health of the Russian Federation
Russian Federation

 post-graduate student  

Russian Federation, 127299, Moscow, st. Priorova 10 



V. V. Konovalov
Federal State Budgetary Institution National Medical Research Center of Traumatology-Orthopedics named after N. N. Priorov of the Ministry of Health of the Russian Federation
Russian Federation

 post-graduate student  

Russian Federation, 127299, Moscow, st. Priorova 10 



References

1. Zagorodny NV, Kolesnik AI, Lazarev AF, Solod EI, Ochkurenko AA, Bukhtin KM, et al. Current trends in the surgical treatment of patients with pelvic and acetabular injuries. The Genius of Orthopedics. 2020; 26 (2): pages 266–274 doi: 10.18019/1028-4427-2020-26-2-266-274

2. Bondarenko AB, Kruglykhin IV, Plotnikov IA, Voitenko AN, Zhmurkov OA. Features of the treatment of pelvic injuries in polytrauma. Polytrauma. 2014; 3: 46–57

3. Donchenko SV, Dubrov VE, Slinyakov LYu, Chernyaev AV, Lebedev AF, Alekseev DV. Algorithm of surgical treatment of unstable pelvic ring injuries. Bulletin of Traumatology and Orthopedics. 2013; 4: 9–16

4. Shapkin YuG, Seliverstov PA. Tactics of treatment of unstable pelvic injuries in polytrauma. Novosti Khirurgii. 2015; 23 (4): 452–459. doi: 10.18484/2305-0047.2015.4.4525

5. Smirnov AA. Surgical treatment of vertically unstable pelvic injuries (type C according to the classification of AO). Traumatology and orthopedics of Russia. 2012; 1 (63): 73–76

6. Kanakaris NK, Angoules AG, Nikolaou VS, Kontakis G, Giannoudis PV. Treatment and outcomes of pelvic malunions andnonunions: a systematic review. Clin Orthop Relat Res. 2009; 467 (8): 2112–24. doi: 10.1007/s11999-009-0712-2

7. Balbachevsky D. Treatment of pelvic fractures – a national survey. Injury. 2014; 45 (5): S46–51. doi: 10.1016/S0020-1383(14)70021-X

8. Bialik EI, Fine AM. Staged treatment of injuries in the area of the pubic articulation using an original fixator in victims with concomitant pelvic trauma. Polytrauma. 2013; 4: 30–34

9. Enninghorst N, Toth K, King KL, McDougall D, Mackenzie S. Acute definitive internal fixation of pelvic ring fractures in polytrauma patients: a feasible option. J Trauma. 2010; 68 (4): 935–41. doi: 10.1097/TA.0b013e3181d27b48

10. Stover MD, Edelstein AI, Matta JM. Chronic Anterior Pelvic Instability: Diagnosis and Management. Journal of the American Academy of Orthopaedic Surgeons. 2017; 25 (7): 509–517 doi: 10.5435/JAAOS-D-15–00338

11. Martel II, Swedes VV. Possibilities of external fixation according to the Ilizarov principles with closed restoration of the shape and stability of the pelvic ring in patients with chronic pelvic injuries. Genius of Orthopedics. 2013; 2: 5–9

12. Kolesnik AI, Dokalin AYu, Solodilov IM, Frolov EB, Angalev EM, Pirogov MA, et al. Retrospective analysis of the results of conservative and surgical treatment of patients with pelvic injuries. Kursk scientific and practical bulletin «Man and his health». 2017; 2. doi: 10.21626/vestnik/2017-2/03


Review

For citations:


Lazarev A.F., Solod E.I., Gudushauri Y.G., Kalinin E.I., Konovalov V.V. Experimental study of the strength characteristics of fixation of the symphysis pubis with an original plate for the reconstruction of the anterior pelvic semi-ring. Acta Biomedica Scientifica. 2021;6(3):185-192. (In Russ.) https://doi.org/10.29413/ABS.2021-6.3.19